TC-3000, TC-3000X SPECIFICATIONS
TC-3000 TC-3000X
Temperature
Temperature range 4°C to 99°C 4°C to 99°C
Temperature set point precision 0.1°C 0.1°C
Block uniformity (over full range) ± 0.1°C ± 0.2°C
Temperature accuracy ± 0.2°C ± 0.2°C
End of program cooling below ambient Yes Yes
Heating/Cooling Rate
Heating rate, between 55°C and 90°C
on a 50°C to 95°C step:
0.2 ml micro tubes 3.6°C/sec 3.3°C/sec
0.5 ml micro tubes 3.6°C/sec 3.3°C/sec
Cooling rate, between 90°C and 55°C
on a 95°C to 50°C step:
0.2 ml micro tubes 2.0°C/sec 2.0°C/sec
0.5 ml micro tubes 1.8°C/sec 2.0°C/sec
Heated Lid
Adjustable heated lid No Yes
Heated lid enable/disable Yes Yes
Selectable heated lid temperature 100°C to 115°C 100°C to 115°C
Over-temperature cut-out Fixed at 145°C Fixed at 145°C
Temperature sensor Thermistor Thermistor
Heater type Etched foil Etched foil
Heater power 33W 33W
Warm up time 4 min approximately 4 min approximately
(ambient to 105°C) (ambient to 105°C)
The heated lid will only come on
if the set temperature is above 35°C
Programming
Program naming using alpha keys Yes Yes
Program password protection Yes Yes
Number of programs (3 step) 80 80
Maximum number of steps per stage 35 35
Maximum number of stages per program 25 25
Minimum hold time 1 sec 1 sec
Maximum hold time 18 hrs 18 hrs
Incremental/decremental temperature Yes Yes
Incremental/decremental step hold time Yes Yes
Programmable ramp rate (heat or cool) Variable in 0.1°C/sec steps Variable in 0.1°C/sec steps
Programmable ramp rate resolution 0.1°C 0.1°C
19